2 The ion with the mass number of 4Easy to identify with other gas by mass spectrometry
3 Small molecular diameterEasy to go through a small hole
4 Non-toxic, non-flammable, and inactive gasSafety use in various fields, environments, and products
5 Small adsorption energy Easily exhaust even if gas enters 6 Non-CFC gas No damage to the global environmentANALYZER:
Analyzer tube.TMP:
Main vacuum exhaust pumpTV1:
Test valve for micro leakTV2:
Test valve for a small leakTV3:
Test valve for big leakTEST PORT:
Test object connection port PG: Pirani gaugeCAL:
Calibrated leakCLV:
Calibrated leak, Vent valveVENT PORT:
Atmospheric pressure return portMagnetic field deflection type mass spectrometry
The current value of the ion collector varies depending on the Helium partial pressure level in a vacuum.
Adoption of a hybrid turbomolecular pump and the effect of a reverse diffusion.
The test port can be connected at higher pressure while maintaining the pressure of the analyzer tube.
* Reverse diffusion phenomenon: It is a phenomenon that a part of a small gas molecule such as Helium, which is difficult to be evacuated, reaches the analyzer tube via a molecular pump.
"Leak reference device" to convert a Helium partial pressure level of an analyzer tube into an amount unit of leakage.
3/sec range.Membrane type. For vacuum method / E-7, E-8, E-9, E-10 Pa・m/sec range.Built Helium gas. Utilizing the Helium phenomenon that transmits the glass. It has a Temperature coefficient and secular attenuation coefficient.
Channel type. Mainly for Sniffing method (Capable for Vacuum Method)/ E-4, E-5, E-6 Pa・m3/sec range.
Helium is supplied from the outside. The vacuum method uses helium at the atmosphere pressure (Differential pressure 0.1MPa). It has a suitable structure for relatively large leakage.
The amount of leakage indicated by the minimum current value detectable by the Leak detector is sensitivity. The smaller the amount is, the better the sensitivity is.
Sensitivity calibration is performed by measuring the calibrated leak regularly.
The smallest leak is expressed as a standard air leak, that can be clearly detected by a leak detector.
A method of detecting Helium leaking that comes to the outside (atmosphere) is by filling Helium into the test object and pressurizing it.
The sniffer unit is used by attaching it to the test port of LD and the vacuum pressure at the test port of LD is maintained at the level that allows continuous operation at TV2. unit.
In addition to a common type that controls the pressure by the orifice at the tip of the probe, there is a sniffer type that is equipped with a pump for suction.